A Master-Slave Robot Manipulator for Needle-Based Teleoperation in MRI Chamber

Percutaneous needle procedures in the liver and kidney require accurate targeting, but inside a closed-bore MRI scanner the strong static field (an intense homogeneous 3T field, per the paper) rules…

The paper’s first page · arXiv:2608.06354
arXiv
2608.06354 · Preprint · not peer-reviewed
Published
2026-08-06 · revised · 2026-08-07
Authors
Omar Curiel, Jing-Yuan Huang, Po-Chih Chen, Ji Ma, Qing Dai, Wenqi Zhou, David Lu, Holden H. Wu, Tsu-Chin Tsao

In plain words

Percutaneous needle procedures in the liver and kidney require accurate targeting, but inside a closed-bore MRI scanner the strong static field (an intense homogeneous 3T field, per the paper) rules out ferromagnetic motors, encoders, and much of the electronics that ordinary robots rely on.

AI summary

Overview

Research area: Medical robotics — MRI-compatible, teleoperated master-slave needle manipulators for image-guided percutaneous abdominal interventions (liver biopsy and ablation).

Technical level: Advanced. The paper assumes familiarity with hydrostatic/fluid transmission, modified Denavit–Hartenberg kinematics, MR safety standards (ASTM F2503), and teleoperation control architectures.

Scope: A table-mounted, MR-safe, 2+1-DoF fluid-actuated needle angulation and insertion robot with a manual master, a motorized digital master, and collaborative control modes, characterized through bench tests, MR scans, and an in-vivo pig experiment.

What This Paper Is About

Percutaneous needle procedures in the liver and kidney require accurate targeting, but inside a closed-bore MRI scanner the strong static field (an intense homogeneous 3T field, per the paper) rules out ferromagnetic motors, encoders, and much of the electronics that ordinary robots rely on. The authors build a robot that stays in the scanner room without disturbing the image, while the surgeon works either at the bedside through a mechanical joystick or from the control room through a motorized digital master. The goal is a device that fits within the roughly 20 cm gap between the patient's skin and the inner wall of a 60 cm bore, delivers needle angulation and insertion with force feedback, and supports manual, digital, hybrid, and collaborative control.

Key Contributions

  1. An MR-safe, table-mounted needle robot for closed-bore scanners. The mechanism is a P-RRR-P serial-parallel hybrid: a 2-DoF angulation stage built on a four-bar linkage riding a semi-circular arch (radius 30 cm), plus a needle insertion axis. It uses a fixed center of motion at a spherical pivot tangent to the pre-insertion skin entry point, instead of active or body-mounted remote-center-of-motion designs.

  2. A purely fluid (hydrostatic) transmission with axis-specific actuator choices. Elastomeric piston cylinders handle angulation because their high impedance and low leakage suit load holding; low-friction graphite/glass piston cylinders handle needle insertion because sub-newton force transparency matters along that axis. Tubing lengths span 15–45 ft (4.6–13.7 m).

  3. A multi-master architecture that sums inputs in the fluid-volume domain. A human-operated master and a motorized digital master act on the same axes simultaneously, which the authors contrast with prior split-axis or mode-switchable hybrid controllers. Table I classifies this paper as the only entry among the surveyed needle robots marked as both force-feedback capable and collaborative (with respiratory motion compensation also listed as "Yes").

  4. Three demonstrated collaborative behaviors plus an in-vivo feasibility result: collaborative tissue penetration, a fault-triggered pressure-regulation virtual fixture, and IMU-based angulation feedback compensation, followed by a bedside manual-mode needle targeting sequence in a live pig.

Main Findings

  • MR compatibility: Using a spin echo sequence (TR/TE = 500/20 ms, flip angle 90°, FOV 250 mm × 250 mm, resolution 0.98 mm × 0.98 mm), SNR differed by 0.11% with the robot at the bedside and 0.59% with it inside the bore. Distortion differed by (0.2, 0.2)% (sagittal, axial) at the bedside and (0.52, 0.2)% in-bore, supporting the MR-safe classification.

  • Workspace: Needle angle spans −40 to 21 degrees in the sagittal plane and −65.5 to 50 degrees in the axial plane relative to gravity, with a 1 mm resolution density map. The reachable volume intersects 33.3–50% of the liver volume under assumptions of a 10 cm skin depth, ~15 cm anteroposterior diameter, and 12 cm needle length. Ten static needle poses scanned in the scanner all fell inside the convex hull of the modeled workspace.

  • Angulation transmission quality: Master-to-slave motion correlation was 0.945 in compression and 0.975 in tension over a 40 mm range, with an identifiable 2.5 mm dead-zone from the elastomeric seals. For a 30 mm peak-to-peak, 5 s trapezoidal waveform, line pressure fluctuated around 5 psi (14.2 N) during compression and settled at 2.3 psi (5.7 N), implying roughly a 6 N holding force and 6–8 N of stiction. Bandwidth was 1 Hz with a 100 ms delay over 40 ft of tubing.

  • Needle-axis transparency: Near one-to-one load-free motion transmission over ±40 mm, with under 5% transmission loss even at 0.2 mm steps. Force transparency exceeds 85% for lines up to 30 ft and loses only 30–50% for master input forces of at least 0.5 N. The insertion servo alone has 4 Hz bandwidth; with 15 ft of tubing the master-to-slave bandwidth is about 2.25 Hz.

  • Collaborative penetration: Against a 4 mm silicone rubber skin phantom, unassisted puncture produced a 1.97 N peak load. Superposing a 20 Hz oscillating input from the motorized master (operator input at 45 mm/s, oscillation swept from 0–20 Hz) reduced needle deflection, average reaction force, and average internal fluid pressure by 25–50%.

  • Fault-driven virtual fixture: When internal line pressure exceeded 3 psi, the motorized masters switched from position to pressure regulation at 0 gauge, absorbing the human master's pressure waves. The recorded digital-master motion was the inverse waveform of the human master's input.

  • Angulation feedback compensation: Using a 100 Hz IMU read-back (in place of MRI-derived angle feedback, which the paper notes can run at 0.5–10 Hz), closing the loop at t = 57 s reduced azimuthal RMS error from 3.81 to 2.95 degrees and elevation RMS error from 2.19 to 1.17 degrees; maximum errors fell from 14.0 to 11.3 degrees and 6.43 to 4.96 degrees respectively.

  • In-vivo feasibility: In a pig experiment, the surgeon inserted the needle roughly 2–2.5 cm through the skin at a fixed angle, then used 2D sagittal image feedback and force cues to retract slightly, adjust the angle in situ when the tip missed the target centroid, and advance to the target — shown at roughly 5–7 cm depth. The paper reports this as preliminary manual-mode validation only; no quantitative targeting accuracy for the animal study is reported.

Methodology in Plain English

The team split the robot into a slave mechanism that sits on the scanner bed and masters that the surgeon operates remotely, connecting them with sealed tubes of fluid rather than wires. Pushing a master piston displaces fluid, which moves the corresponding slave piston; because the fluid path carries no electricity, no motors or encoders need to be near the magnet. Two fluid circuits drive the angulation axes through a four-bar linkage, and a separate circuit drives the needle in and out.

To make the fluid links behave usefully, the authors chose different seals per axis: rubbery elastomeric seals hold position well but add friction and a small dead-zone (partly mitigated by dithering), while graphite pistons on a glass cylinder give low-friction, nearly one-to-one force and motion transfer for insertion. On top of the human master, they added a motorized master whose fluid output merges with the human's, so software can add oscillations, cancel motion, or take over pressure regulation when a fault is detected. Validation proceeded in layers: SNR and distortion measurements in the scanner, pose checks against the kinematic workspace model, bench characterization with rack-and-pinion and load-cell rigs, closed-loop error statistics, and finally a live pig procedure in bedside manual mode.

Why This Matters

The paper targets a practical gap: many MRI-guided needle robots are conditionally safe rather than fully MR safe, rely on piezoelectric or ultrasonic motors that can inject noise into the image, or are body-mounted and therefore drift with the patient's diaphragm during breathing. A table-mounted, fixed-center-of-motion device with no in-bore electronics sidesteps several of these failure modes at once, and the multi-master scheme lets a surgeon keep hands-on control while software assists.

Real-world applications:

  • Liver tumor biopsy and ablation: reaching hepatic lesions at a mean depth of about 5 cm (up to 15 cm) while avoiding vessels, ribs, and adjacent organs.
  • Renal and abdominal percutaneous interventions requiring oblique entry angles of 5 degrees or more from the axial/transverse plane.
  • Assisted penetration of stiff tissue, where superposed high-frequency motion reduced puncture forces and deflection.
  • Motion-compensated targeting during procedures that exceed a breath-hold, using feedback to correct orientation and position errors.

Industry relevance: the design points toward a low-cost, disposable-friendly fluid transmission path for MR-safe instrumentation, useful to developers of interventional MRI suites, needle-guidance robot startups, and surgical planning software that needs a robot API exposing manual, digital, and shared control.

Future Directions

  • Closed-loop MRI-guided needle placement using the same architecture, with image-derived rather than IMU-derived needle pose feedback.
  • Adaptive motion compensation for interventions that run longer than a breath-hold, exploiting spatio-temporal needle-tissue mechanical interaction models.
  • Remote tissue impedance estimation built on the force fidelity of the low-friction glass actuator, offered as an additional interpretable signal to the surgeon.
  • Scaling the mechanism from the 60 cm bore and 40–70 kg pig subjects to human 70 cm bores, plus quantitative clinical validation beyond the single preliminary in-vivo case.

Target Audience

Robotics researchers working on medical and MRI-compatible manipulators; graduate students in mechanism design, fluid power, or teleoperation control; and clinicians or industry engineers involved in interventional MRI, needle guidance, and surgical robot translation who want a concrete example of how MR safety constraints reshape actuator, sensor, and control choices.

Authors’ abstract

We present a MR safe, master-slave robot manipulator for abdominal interventions in the MRI chamber. A human operated 2+1-DoF master controller manipulator transmits motion and force to a 2+1-DoF slave manipulator via fluid transmission. Jointly, a digital master controller provides multimodal control capability beyond common split axis or mode switchable hybrid human-digital controller configurations found in previous studies. High input impedance, low-leakage, elastomeric…