# An Open-Source da Vinci X/Xi Instrument Module for Laparoscopic Surgery Research

Robotics.world

> Research robots that study electrosurgical tasks need to control both how an instrument moves and when energy fires, and to keep a record of which of those came from the surgeon and which from the…

**arXiv**: 2611.21771 · Preprint · not peer-reviewed

**Published**: 2026-11-16

**Authors**: Douglas McDonough, Justin Liao, Jiawei Ge, Antony Goldenberg, Yvonne Zhang, Mackenzie A Eagleson, Ravi Prakash, Axel Krieger, Chunwoo Kim

## In plain words

Research robots that study electrosurgical tasks need to control both how an instrument moves and when energy fires, and to keep a record of which of those came from the surgeon and which from the software.

## AI summary

## Overview

**Research area:** Surgical robotics — hardware and firmware for research platforms that actuate da Vinci X/Xi laparoscopic instruments and control/record electrosurgical energy.

**Technical level:** Advanced. The paper assumes familiarity with robot manipulators (UR5/UR5e), motor actuation, coupling matrices and kinematics, relay circuits, ROS 2, and electrosurgical generators.

**Scope (one sentence):** The paper describes an open-source five-motor instrument driver that mounts da Vinci X/Xi instruments on a UR5/UR5e arm, together with a relay-based interposer that lets software command monopolar and bipolar energy while giving the surgeon's foot pedals priority and logging the source of every activation.

## What This Paper Is About

Research robots that study electrosurgical tasks need to control both how an instrument moves and when energy fires, and to keep a record of which of those came from the surgeon and which from the software. Existing open adapters for da Vinci instruments actuate only simple articulated tools such as graspers and scissors, and none of them provides a documented interface for remote software activation and logging of electrosurgical instruments. This paper presents a compact, open-source module that drives wristed instruments and the SureForm 45 Stapler, and adds an electrosurgical interface that lets software request energy while physical pedals always win, recording the source of each request alongside robot motion and video.

## Key Contributions

1. **A five-motor instrument driver for da Vinci X/Xi tools** built around five DYNAMIXEL XC330-M288-T servos, transmitting torque through timing belts (22-tooth pulleys, 106 mm and 128 mm belts) to spring-loaded drive discs, mounted to a UR5/UR5e through an ISO 9409-1-50-4-M6 flange. It accommodates wristed instruments and, uniquely among the adapters the authors cite, the SureForm 45 Stapler.
2. **An electrosurgical interposer (ESU-TAP v2)** that sits between the foot pedals and the generator, supports software activation of monopolar CUT, monopolar COAG, and bipolar modes, passes the pedals directly to the generator when unpowered, and records the state and source (pedal, software, or takeover) of each activation request.
3. **Calibrated instrument kinematics**, including a shared coupling matrix for the fenestrated bipolar forceps, monopolar curved scissors, and large needle driver, plus a separate matrix for the SureForm 45 stapler, with an experimental calibration procedure that improved measured-to-commanded rotation ratios.
4. **An open release** of CAD, firmware, and ROS 2 driver code in the ARTS repository, with evaluation on the open-source three-arm ARTS platform covering bench tests of the energy interface, ex vivo electrosurgical tasks, instrument exchange time, kinematic accuracy, and homing repeatability.

## Main Findings

- **Energy interface works in all three modes:** Monopolar CUT, monopolar COAG, and bipolar activation succeeded through both software commands and physical pedals, confirmed by the DRE ASG-300's audible tones and visual indicators. All three modes remained operable through the pedals with the interposer unpowered, confirming the hardware bypass.
- **Software timeout matched the design value:** The board-reported software activation timeout was **503.1 ± 3.8 ms** across eight trials, consistent with the programmed 500 ms timeout.
- **Fast pedal takeover:** During a takeover test, the board reported software-controlled CUT released and bipolar closed **11 ms** later, and subsequent software commands were rejected until software control was explicitly re-enabled.
- **Communication loss fails safe:** Terminating the micro-ROS agent during an active software command caused the generator to cease activation without additional commands; on reconnection the interposer reported software control disabled and all outputs deactivated.
- **Complete activation logging:** The interposer recorded all **12** activation and deactivation events from **six** foot pedal presses, preserving the correct sequence and identifying the physical pedals as the source.
- **Faster instrument exchange than DASH Xi:** Across **20** trials, instrument exchange took **2.41 ± 0.34 s**, compared with **3.39 ± 0.51 s** reported for the DASH Xi adapter (measured following the same protocol, excluding automatic homing).
- **Module mass:** Estimated mass ranged from **0.740 kg** with no instrument to **1.270 kg** with the SureForm 45 Stapler. Other configurations: large needle driver 0.850 kg, fenestrated bipolar forceps 0.830 kg, permanent cautery hook 0.960 kg, monopolar curved scissors 1.030 kg.
- **Calibration improved pitch and yaw accuracy:** Initial coupling coefficients were [1.0, 0.95, 1.28, 0.63] and measured-to-commanded rotation ratios were **0.993, 0.904, 1.14** for roll, pitch, and yaw, with roll-yaw and pitch-yaw axis angles of **80.1°** and **87.9°**. After fine-tuning to [1.0, 1.050, 1.12, 0.57], the ratios became **0.99, 1.00, 1.02** and the axis angles improved to **89.85°** and **89.3°**. Hysteresis inherent to cable-driven mechanisms remained visible in the plots.
- **Repeatable current-based homing:** Mean homing time was **34.9 ± 0.2 s** over ten trials with the Permanent Cautery Hook. Encoder spread ranged from SD 2.6 ticks (0.24°) on pitch to SD 4.4 ticks (0.36°) on yaw; pitch had the smallest range at 9 ticks (0.84° equivalent joint angle) and yaw the largest at 14 ticks (1.12°).
- **Ex vivo tasks succeeded:** Video review showed suction clearing simulated blood and monopolar coagulation at a splenic wound, splenic wound edges remaining approximated after the bipolar forceps jaws opened, and tissue separation during monopolar gallbladder dissection.

## Methodology in Plain English

The team started from an existing open adapter (MOPS) and redesigned it around smaller, lighter servo motors. Shrinking the actuators from the MX-28T to the XC330-M288-T cut the bounding-box volume by 72% and the mass per actuator from 77 g to 23 g, which freed enough room to add a fifth motor. Motors sit on the outside of the adapter and drive the instrument's input discs through belts, so the package stays compact. Each drive shaft holds a small spring plunger with a 3 mm stroke: when an instrument is pushed in, the springs compress and then push the adapter discs into the instrument's recesses once their prongs line up, which makes engagement simple. Snap-fit clips hold the instrument, with optional star knobs for extra security and a cannula-like extension to reduce shaft wobble.

To convert motor rotations into instrument joint motion, the team uses coupling matrices — one shared by the forceps, scissors, and needle driver, and a separate one for the stapler. They calibrated the coefficients by manually articulating the tool, measuring joint angles and the corresponding disc rotations, then refining the numbers and repeating the test with an optical marker on the forceps, commanding 5° increments from −40° to 40° in roll and pitch and −25° to 40° in yaw. For homing, a current-based routine rotates joints until motor current rises at mechanical limits and sets zero at the midpoint, mirroring the da Vinci Xi auto-homing idea; joints without clear limits are aligned visually.

For energy, they built a small circuit board between the pedals and the generator using six relays — three for activation (one per mode) and three bypass relays that keep the pedals hard-wired to the generator when the board is unpowered. An ESP32 running micro-ROS samples the pedals on a 10 ms control task with 30 ms debounce, so pedal forwarding works even without the host. Software activation must be explicitly armed, must be renewed every 100 ms, and expires 500 ms after the last renewal; any pedal press cancels software control and hands authority back to the surgeon. They then tested timeout behavior over eight trials, pedal takeover, simulated communication loss, and event recording, and ran three ex vivo tasks on the ARTS platform — suction plus monopolar coagulation on a splenic wound, bipolar coagulation of wound edges, and monopolar cutting during gallbladder dissection — while recording video, kinematics, and activation events.

## Why This Matters

**Impact on research:** The module lowers the barrier to studying energy-delivering surgical tasks on a research robot. Because activation events are labeled by source and timestamped alongside motion and video, datasets gain the action labels that learning-from-demonstration methods need — something the authors state was missing from prior electrosurgical data-collection work. The design also extends open research hardware beyond simple graspers to a stapler-class instrument.

**Real-world applications:**

- **Surgical training and skill assessment** — logged pedal and instrument motion gives an objective record of when and how energy was used during a task.
- **Learning from demonstration and autonomy** — labeled energy events paired with kinematics support imitation learning and automated dissection research.
- **Electrosurgical safety studies** — the pedal-priority and lease-timeout design gives a concrete, testable model for how software-commanded energy should yield to the surgeon.
- **Bench and ex vivo protocol development** — a reproducible rig for coagulation, cutting, and instrument-exchange studies before animal or clinical work.

**Industry relevance:** The interposer's architecture — hardware bypass when unpowered, explicit arming, 100 ms renewals with a 500 ms expiry, and pedal takeover — is a template for how third-party software might safely request energy on a commercial generator without modifying it or carrying its RF output. The 2.41 s instrument exchange and sub-1.3 kg payloads are also practical numbers for groups integrating surgical instruments onto general-purpose arms.

## Future Directions

1. **Staple firing.** The coupling matrix for the SureForm 45 Stapler's shaft roll, wrist articulation, and jaw motion is defined, but firing is described as a separate development step to be evaluated in future work.
2. **Automatic instrument identification.** The module requires manual selection of the instrument configuration, whereas DASH uses RFID identification to load instrument-specific kinematics and zero automatically — a clear gap to close.
3. **Broader generator and interposer validation.** The authors note that other generators require a compatible footswitch interface and verification of wiring and signal levels, and that the reported state reflects commanded relay outputs without contact or RF feedback; independent timing of contact opening and RF output is not reported.
4. **Fault coverage.** A hung controller or a welded relay is not handled by the board and falls to the generator's power switch — a safety gap the current design leaves open.

## Target Audience

Robotics researchers building or extending surgical research platforms, particularly groups using the dVRK, MOPS, DASH, or ARTS and wanting to add energy-delivering instruments and labeled data collection. It also suits engineers working on human-in-the-loop safety interfaces where software commands must defer to a physical operator, and surgical researchers who need a reproducible bench setup for electrosurgical studies. Readers without a background in mechanism design, motor control, or ROS 2 will find the hardware and firmware sections dense.

## Authors’ abstract

Studying electrosurgical tasks on a research robot requires control of both instrument motion and energy activation, along with a record of when each action occurs. This paper presents an open-source module for da Vinci X/Xi instruments on a UR5/UR5e robotic arm. Five motors transmit torque through timing belts to spring-loaded drive discs. An electrosurgical interposer accepts software commands, gives the surgeon's foot pedals priority, and records the source of each…

[The full abstract on arXiv](https://arxiv.org/abs/2611.21771)

[Read the original paper](https://arxiv.org/abs/2611.21771)

## Related papers

- [ARTS: An Open-Source, Reproducible Three-Arm Platform for Autonomy Research in Minimally Invasive Surgery](https://robotics.world/paper/arts-an-open-source-reproducible-three-arm-platform-for-autonomy-2611.15135) 2026-11-11 · cs.RO · Also by Antony Goldenberg, Axel Krieger +4
- [Supervised Autonomous Bimanual Suction and Cauterization for Bleeding Control in Low Grade Splenic Trauma](https://robotics.world/paper/supervised-autonomous-bimanual-suction-and-cauterization-for-bleeding-2611.00237) 2026-10-30 · cs.RO · Also by Axel Krieger, Jiawei Ge +2
- [Velocity Optimization via Thermal Imaging Enhances Autonomous Tumor Resection](https://robotics.world/paper/velocity-optimization-via-thermal-imaging-enhances-autonomous-tumor-2610.27562) 2026-10-20 · cs.RO · Also by Axel Krieger, Jiawei Ge
- [Occupancy Network-Guided Autonomous Robotic Partial Nephrectomy](https://robotics.world/paper/occupancy-network-guided-autonomous-robotic-partial-nephrectomy-2609.16186) 2026-09-14 · cs.RO · Also by Axel Krieger, Jiawei Ge
- [An Open-Source ROS 2 Framework for Shared Control in Surgical Teleoperation: Integrating the Haply Inverse3 with the da Vinci Research Kit](https://robotics.world/paper/an-open-source-ros-2-framework-for-shared-control-in-surgical-2611.26485) 2026-11-19 · cs.RO · Key terms of the title in common

---

Canonical: https://robotics.world/paper/an-open-source-da-vinci-x-xi-instrument-module-for-laparoscopic-surgery-2611.21771