Skip to content
Robotics.world

Research

An Open-Source ROS 2 Framework for Shared Control in Surgical Teleoperation: Integrating the Haply Inverse3 with the da Vinci Research Kit

Robotic surgery separates the surgeon from the patient, so touch is lost and tissue forces must be inferred visually. The few commercial platforms that now restore force feedback are closed, which…

arXiv
2611.26485 · Preprint · not peer-reviewed
Published
2026-11-19
Authors
Balint Varga, Sebastian Fritz, Gergely Kiss, Kristof Takacs, Peter Galambos

In plain words

Robotic surgery separates the surgeon from the patient, so touch is lost and tissue forces must be inferred visually. The few commercial platforms that now restore force feedback are closed, which blocks open research on how to design haptic guidance (not just force reflection).

AI summary

Overview

Research area: Robot-assisted surgery / surgical teleoperation, haptics, and haptic shared control; human–robot interaction on the da Vinci Research Kit (dVRK).

Technical level: Advanced — the paper assumes familiarity with ROS 2 middleware, kinematic frames and coordinate transforms, virtual fixtures, admittance/impedance-style force rendering, and within-subjects experimental design with paired statistics.

Scope (one sentence): The paper presents and validates a ROS 2-native, open-source software framework that lets a compact 3-DOF Haply Inverse3 haptic device drive the dVRK, together with two simple virtual-fixture-inspired guidance modes tested in a 20-participant simulated hot-wire user study and demonstrated on physical hardware.

What This Paper Is About

Robotic surgery separates the surgeon from the patient, so touch is lost and tissue forces must be inferred visually. The few commercial platforms that now restore force feedback are closed, which blocks open research on how to design haptic guidance (not just force reflection). The authors build an open ROS 2 framework that couples a small, low-cost 3-DOF haptic device to the dVRK, define a mapping that turns its 3-DOF translation plus separately tracked stylus orientation into 6-DOF instrument commands, and use two guidance modes as reference implementations to check that the whole chain works with real users and on real hardware.

Key Contributions

  1. An open-source ROS 2 architecture for Haply Inverse3 + VerseGrip ↔ dVRK teleoperation, including a fully specified mapping pipeline, a driver wrapping the manufacturer's WebSocket-based haply-inverse-service API, custom ROS 2 message definitions, RViz visualization, demo nodes, and safety mechanisms (per-axis force saturation, a command watchdog with τ_wd = 0.5 s, and a connection-loss shutdown after τ_disc = 50 s). Released under the MIT License at https://github.com/ABC-iRobotics/haply_ros2_interface.
  2. A relative (incremental) mapping strategy combining 3-DOF translational device input with VerseGrip orientation into 6-DOF patient-side manipulator (PSM1) commands, with a Cartesian variant and a joint-space variant (joint-space used in all experiments), plus clutching on VerseGrip button B and gripper toggling on button A.
  3. Two virtual-fixture-inspired guidance modes as reference implementations — M1 (linear stiffness, k = 80 N/m, b = 2 N·s/m) and M2 (progressive piecewise-linear stiffness, k₁ = 40 N/m and k₂ = 120 N/m, same damping), analytically constructed so both render identical maximum force at the outer zone boundary (d_o = 0.015 m).
  4. An empirical human-in-the-loop evaluation on a teleoperated hot-wire benchmark in simulation with N = 20 participants, plus a demonstration of the same architecture on physical dVRK hardware.

Main Findings

  • Contacts with the wire (M1 significant): mean contacts were 31.85 (SD 16.78) for manual M0, 20.90 (SD 12.86) for linear guidance M1, and 24.35 (SD 15.69) for progressive-stiffness M2. M0 vs. M1: t(19) = 3.18, p = .005, d = 0.71 — significant at the Bonferroni-corrected α = .025.
  • Contacts with the wire (M2 not significant): M0 vs. M2 gave t(19) = 1.58, p = .130, d = 0.35, with a mean reduction of 7.50 contacts — a trend in the same direction but not statistically significant.
  • Task completion time (M1 significant): mean times were 96.31 s (SD 37.47) for M0, 75.01 s (SD 33.06) for M1, and 84.25 s (SD 36.26) for M2. M0 vs. M1: t(19) = 3.45, p = .003, d = 0.77, a mean reduction of 21.30 s.
  • Task completion time (M2 not significant): M0 vs. M2 gave t(19) = 1.74, p = .098, d = 0.39.
  • Subjective ratings trended favorably but none survived correction: on a seven-point scale from −3 to +3, perceived demand (Q1) had medians +2.0 (M0) vs. +1.0 (both M1 and M2); sense of agency (Q2) had medians −1.0 (M0), +1.0 (M1), +0.5 (M2); satisfaction (Q3) had medians −1.0 (M0) vs. 0.0 (both M1 and M2). Nominal significance before correction appeared for Q1 M0 vs. M1 (W = 15.5, p = .034, r = .659) and Q3 M0 vs. M1 (W = 38.5, p = .038, r = .550); effect sizes across all comparisons ranged from r = .450 to r = .659 (medium-to-large).
  • Simpler guidance was more robust for inexperienced operators: the authors conclude that linear feedback (M1) may be preferable to the more complex progressive profile (M2), and attribute the non-significant M2 results partly to limited statistical power at N = 20.
  • The framework rendered forces as designed: a representative Mode 1 trial shows the Haply Inverse3 producing the intended guidance forces, supporting the interpretation that subjective ratings reflect operator perception rather than system limitations.
  • Simulation fidelity: the simulated study reused the entire software stack of the physical system; only the physical PSM (replaced by a kinematic model) and the endoscopic camera (replaced by an RViz-rendered virtual hot-wire environment) differed.
  • Hardware transfer: the architecture is reported to be demonstrated on physical dVRK hardware, though the excerpt provided here is truncated before Section VII and the quantitative hardware results are not reported in the available content.

Methodology in Plain English

The team wrote a ROS 2 driver that talks to Haply's WebSocket service, so the Inverse3's position, velocity, VerseGrip orientation, and buttons arrive as ordinary ROS 2 topics, and force or position commands go back out (position mode is implemented as a software PID loop that converts desired position into corrective force). Because the Inverse3 only senses three translational degrees of freedom, the bridge node records a reference snapshot — Haply position, stylus roll-pitch-yaw, and the current robot state — and thereafter sends only increments: scaled position changes (s_p = 0.25) and scaled, wrapped orientation changes, with a fixed axis-flip matrix diag(−1, 1, −1) to align the Haply frame to the PSM base frame. In the joint-space variant used in all experiments, the increments are mapped onto the six PSM joints (outer yaw, outer pitch, insertion, instrument roll, wrist pitch, wrist yaw) with a gain g = 10 rad/m chosen so lateral and axial motion feel similarly scaled at roughly 0.1 m insertion depth. A button press re-zeroes translation so the user can re-center the device without moving the instrument.

For the benchmark, the user drives a ring along a five-segment wire (total length 0.21 m; segments of 0.04, 0.03, 0.05, 0.05, and 0.04 m; ring diameter 0.02 m). Inside a virtual-fixture zone around the wire (inner radius d_i = 0, outer radius d_o = 0.015 m), the framework pushes the hand back toward the wire center along the unit normal, plus a damping term to kill oscillation. Mode 1 uses one constant stiffness; Mode 2 splits the zone at its midpoint, using softer stiffness inside and stiffer outside, and the outer stiffness is solved analytically so both modes cap at the same force. Because the Inverse3 produces forces between 0 N and 1.2 N, those stiffness values were chosen to stay within range.

Twenty volunteers (7 female, 13 male, ages 25–65, mixed teleoperation experience) each did the task under M0, M1, and M2, with order counterbalanced by a balanced Latin square and five minutes of training per mode (15 minutes total). Only a standard 2D monitor was used, with contact shown by a color change of the wire. Objective measures were contact count and completion time, analyzed with paired t-tests after Shapiro–Wilk checks, with the significance threshold Bonferroni-adjusted to α = .025 for the two planned comparisons. Three Likert questions (demand, agency, satisfaction) were analyzed with Wilcoxon signed-rank tests at the same corrected threshold. The protocol was approved under ethics approval no. 596, with written informed consent and anonymized data.

Why This Matters

Impact on research: The work removes a practical barrier — previously, coupling an external haptic master to the dVRK meant bespoke, legacy-middleware integrations, and published dVRK haptic studies cluster around the sigma.7 and PHANToM family. A modular ROS 2 integration, published under MIT, gives labs a reproducible base for studying haptic shared control rather than only force reflection, and it introduces the hot-wire task as a benchmark for surgical haptic shared control, which the authors note had not previously been used that way.

Real-world applications:

  • Surgical training simulators that give trainees corrective force cues while they learn path-following and force regulation.
  • Low-cost research and educational teleoperation rigs, since a compact 3-DOF device plus orientation tracking substitutes for a much larger 6-DOF haptic master.
  • Shared-control assistance for delicate sub-tasks such as suturing along a planned path, where forbidden-region or guidance fixtures constrain motion near critical anatomy.
  • Bench-marking and comparison studies across labs, enabled by an open stack and a standardized task with a binary error metric.

Industry relevance: The paper's central complaint — that even da Vinci 5, which introduces force feedback, remains closed — is a direct argument for open research platforms. Device vendors and surgical robotics groups get a template for wrapping proprietary device APIs into ROS 2, a demonstration that consumer-grade haptics can serve clinical-grade robots, and evidence that the simplest guidance law may be the most usable one for non-expert operators, which matters for product decisions about how much assistance to layer on.

Future Directions

  • Investigate why progressive stiffness underperformed: the paper reports trends but does not explain them; whether the softer inner region (k₁ = 40 N/m) or the stiffness transition itself is responsible is an open question, as is whether different d_m or k₂ choices would change the outcome.
  • Replicate with more participants and richer instruments: the authors attribute the non-significant subjective results to limited power at N = 20; larger studies, and validated instruments such as NASA-TLX or the System Usability Scale (used elsewhere in the literature they review), would test whether the descriptive trends hold.
  • Quantify performance on physical hardware: the architecture is demonstrated on the physical dVRK, but the provided content does not report hardware performance numbers; a full hardware user study including PSM dynamics and a real endoscopic camera remains to be done.
  • Broaden the device and task space: the framework is demonstrated on a 3-DOF device, so extending to devices with more degrees of freedom, adding force sensing or sensorless force estimation on the patient side, and testing tasks beyond the hot-wire benchmark (e.g., palpation or peg transfer) are natural next steps.

Target Audience

Robotics researchers and graduate students working on surgical teleoperation, haptic shared control, or virtual fixtures; dVRK users who want to swap in a compact, inexpensive haptic master without writing their own middleware; ROS 2 developers interested in a worked example of wrapping a proprietary device API into a real-time control loop; and human-factors researchers looking for a standardized benchmark (the hot-wire task) with a binary error metric for studying guidance and workload.

Authors’ abstract

Surgical teleoperation deprives surgeons of touch, forcing visual inference of tissue forces; the commercial platforms now restoring force feedback remain closed, preventing open research on haptic shared control. This paper provides a system contribution: it presents an open-source, ROS 2-native framework that integrates the compact, three-degree-of-freedom (3-DOF) Haply Inverse3 haptic device with the da~Vinci Research Kit (dVRK). The framework includes a novel mapping…

The full abstract on arXiv
Read the original paper