Integrated Respiratory Sensing and Gated Control for Autonomous Needle Insertion Via CT Bore Robotic Arm
CT-guided lung biopsy currently requires a clinician to step in and out of the scan room for repeated manual needle adjustments while the patient holds their breath, which lengthens procedures and…

- arXiv
- 2610.30452 · Preprint · not peer-reviewed
- Published
- 2026-10-21
- Authors
- Emma Huang, Moira Bohley, Krishna Jamakhandi, Rebecca Fellbaum, Sophia Tang, Sean Tutton, Albert Hsiao, Alexander Norbash, Changwei Chen, Derek Chen, Ananya Rao, Lizzie Peiros, Michael C. Yip
In plain words
CT-guided lung biopsy currently requires a clinician to step in and out of the scan room for repeated manual needle adjustments while the patient holds their breath, which lengthens procedures and increases radiation dose. Breath-holding is also unreliable for the older patients with severe pulmonary disease who most often need the procedure.
AI summary
Overview
Research area: Medical robotics — specifically image-guided percutaneous intervention, combining respiratory motion sensing, state estimation, and gated motion control for a needle-insertion robot that operates inside a CT scanner bore.
Technical level: Advanced. The paper assumes familiarity with Extended Kalman Filters, harmonic signal modeling, lead-lag compensator design, and second-order system identification, though the clinical motivation and experimental results are described in accessible terms.
Scope (one sentence): The paper presents and bench-tests an integrated sensing-and-control stack — contact sensor, harmonic breathing model, EKF-based forecasting, and a four-stage procedural state machine — that lets a CT-bore robot arm insert a biopsy needle only during the exhalation phase of a patient's natural breathing cycle, with anomaly detection to abort unsafe insertions.
What This Paper Is About
CT-guided lung biopsy currently requires a clinician to step in and out of the scan room for repeated manual needle adjustments while the patient holds their breath, which lengthens procedures and increases radiation dose. Breath-holding is also unreliable for the older patients with severe pulmonary disease who most often need the procedure. The goal of this work is to let a robot arm, mounted on an existing CT-bore biopsy robot, sense the patient's quiet breathing through physical contact, predict where the chest will be a few hundred milliseconds in the future, and autonomously drive a needle forward only during the stable exhalation basin — no breath-holding required.
Key Contributions
-
A quiet-breathing motion model. The authors recorded chest motion from eight volunteers using a 3 × 3 grid of nine OptiTrack markers over 10-minute sessions and characterized breath duration, displacement amplitude, and inhalation/exhalation asymmetry, then used those measurements to derive concrete sensing design targets (3.74–6.22 mm amplitude range, 0.37 mm spatial resolution, 5 ms maximum response time).
-
An integrated sensing and control mechanism that works inside the CT bore. A contact sensor converts linear chest displacement into rotation read by a Hall effect sensor, paired with a ToF sensor for the approach phase; a CT-compatible variant uses a clutch-unclutch needle gripper driven by a Dyneema string in a Bowden cable to avoid the metal that would cause CT artifacts.
-
An autonomous four-stage linear needle insertion procedure with gated control. The sequence runs torso approach and contact, breathing cycle sensing, initial drive and dermal puncture, and gated insertion to target depth, with the EKF forecasting chest position at a computed delay horizon and a lead compensator reducing servo tracking lag.
-
Anomaly detection during the autonomous procedure. A normalized innovation squared (NIS) statistic, thresholded against each run's own clean baseline, flags deviations from normal breathing and triggers state-specific safe responses (stop, or go slack).
Main Findings
-
Measured breathing matched literature in magnitude but was faster and more symmetric in the study cohort. Across eight volunteers, breath duration was 4.03 ± 0.97 s (range 2.41–6.44 s) versus a reference value of 4.335 s; displacement amplitude was 4.98 ± 0.62 mm (range 3.94–5.72 mm) versus a reference sternum displacement of 4.33 mm; and the inhalation/exhalation ratio was 0.81 ± 0.12 (range 0.51–1.08) versus a reference of 0.64.
-
The contact sensor exceeded the resolution requirement by a wide margin. The ToF sensor showed up to 8 mm peak-to-peak noise with 2 mm standard deviation, while the contact sensor showed 0.011 mm peak-to-peak and 0.002 mm standard deviation — well below the 0.37 mm design target. The ToF sensor was calibrated at 10 stops between 0 mm and 10 mm, yielding a 2.6 mm offset and 1.05 slope.
-
The EKF forecast beat raw sensor readings. After tuning the process-noise scale to 0.5 and the velocity bound to 10% over 10 trials across five patient profiles, the calibrated parameters gave a mean forecast RMSE of 0.293 mm (max 0.591 mm). On five held-out patient profiles, the forecast achieved 0.165 mm mean RMSE versus 0.235 mm for the raw sensor, a 29.8% error reduction, and outperformed the raw signal on 10/10 validation profiles.
-
The approach-to-contact transition was reliable. In 10/10 trials across five patient profiles, the needle moved from randomized starting distances of 30 to 40 mm to a 1 mm standoff with ±0.3 mm tolerance and no tissue damage.
-
Full autonomous insertions reached target depth in every trial. All five end-to-end trials reached the 40 mm target depth without violating gating parameters, with an average time to full insertion of 14.41 s. The initial drive inserted an average of 17.91 mm; one trial needed two additional gated insertions and the remaining four needed three.
-
Needle-drive dynamics were consistent with and without tissue. Ten free-air step-velocity trials gave K_p = 0.92, ω_n = 27.0 rad/s, and ζ = 0.35, and repeating the trials with the needle inserted into tissue produced no significant difference.
-
Abrupt breathing anomalies were caught at the speed of the sensing lag. Injected disturbances (an amplitude step from 2.0 to 3.0 mm, a rate step from 15 to 20 bpm, both together, and a bounded ±4 bpm wander about 15 bpm with 8 s decorrelation) were detected in 0.17–0.23 s for the three step cases, essentially matching the 0.22–0.25 s sensing lag alone. The slow rate wander took 2.32 s to detect, about 2.1 s of which was genuine detector lag — an intentional design choice since variability of that size occurs in normal breathing.
-
Measured system latency was substantial and is the core problem the forecasting solves. Normalized cross-correlation gave approximately 300 ms of latency on the testbed and roughly 330 ms on the CT bore robot mechanism, with contact mechanics accounting for most of it and the contact sensor running at 100 Hz.
-
Torque sensing showed promise but was not deployable. Torque rose with insertion depth and showed oscillating drops after the needle punctured the back of the phantom, across six trials at three angles and two speeds, but the authors report it is currently limited by motor hardware on the CT bore robot.
Methodology in Plain English
The team first needed to know what normal breathing actually looks like, so they stuck reflective markers on volunteers' chests and tracked them optically for 10 minutes each, extracting the front-to-back component of chest motion. They also captured coughing, hiccups, and talking to see what abnormal signals look like. From those numbers they derived how precise a sensor would need to be.
They then built two test rigs. The first is a bench-top system: a silicone skin phantom on a motorized linear rail that plays back recorded or synthetic breathing profiles, plus a needle-drive mechanism mirroring the robot's. The second is a compacted, metal-minimized version meant to bolt onto the actual CT-bore robot. Sensing works in two stages — a time-of-flight distance sensor for the long approach, then a physical contact probe that rides on the chest and converts displacement into a rotation a Hall effect sensor can read. A threshold decides when to hand off from one sensor to the other rather than fusing them, because their ranges and accuracies differ so much.
Because the chest keeps moving during the roughly 300 ms it takes for a reading to become an action, the team modeled breathing as a sum of a few harmonics with a slowly drifting phase, fit it to a 60-second calibration window using a Fast Fourier transform for the initial rate and ordinary least squares for amplitudes and phases, and then ran an Extended Kalman Filter online to keep refining the model. The filter is stepped forward in time by the computed delay horizon to produce a prediction of where the chest will be when the needle actually moves. Two safeguards matter: the breathing rate state is clamped within a fixed fraction of the initial estimate, because otherwise the filter can silently compensate for a wrong rate by drifting a phase term and still look accurate; and the normalized innovation squared is monitored as a health check, since it should hover near 1 for a well-specified filter.
The needle servo itself is identified as a second-order velocity plant with an extra integrator from position, and a lead compensator is designed by constrained search to keep phase margin adequate, crossover below the plant's natural frequency, and closed-loop lag at the nominal breathing rate as small as possible.
Everything is orchestrated by a four-state machine: approach and contact using the ToF sensor, sense breathing until the filter has converged and passed its consistency and accuracy checks, puncture the skin during an exhalation basin to seat the needle, then drive in increments during successive exhalation basins until the target depth is reached. Between gated pushes the needle is deliberately left compliant so it floats with the chest. If the NIS detector fires, the state machine responds according to its current phase.
Why This Matters
Impact on research: The paper closes a specific gap in CT-bore robotic biopsy — prior systems from Schreiber et al. and Zhang et al. could reach and drive a needle inside the bore but had no mechanism to compensate for breathing, while respiratory-gated systems from Li et al. and Wen et al. still depended on a surgeon to complete the insertion. This work combines both capabilities and, unusually, publishes the estimator design details (process-noise calibration, rate-bounding, NIS diagnostics) that determine whether gated control actually works. The anomaly-detection results are also notable: they show that a filter can appear to track correctly while forecasting at a wrong rate, which is a failure mode other motion-compensation systems may share.
Real-world applications:
- Percutaneous lung biopsy — the direct target, where hitting a moving nodule is the central difficulty and where breath-holding protocols fail for the patients who need the procedure most.
- Patients with severe pulmonary disease — the population for whom breath-holding is impractical, who could be treated without requiring respiratory cooperation.
- Other CT-guided percutaneous procedures — any needle task in the chest or upper abdomen where target motion follows respiration, such as ablation or drainage placement.
- Radiation dose reduction — fewer confirmation scans and fewer clinician trips in and out of the room translate to less exposure for patient and staff.
Industry relevance: The work builds directly on an existing CT-guided needle biopsy robot platform rather than proposing a new device, and it deliberately minimizes metal in the CT-integrated sensor and uses a cable-driven gripper to avoid image artifacts. That makes it a plausible integration path for surgical robotics companies and interventional radiology groups already working inside scanner bores, where the practical bottleneck is workflow time and scanner occupancy rather than raw mechanical capability.
Future Directions
- A more realistic torso phantom. The authors attribute inconsistent passive needle motion during inhalation partly to a phantom that deformed too much, and a needle mechanism that offered high resistance; better skin deformation properties are called out explicitly.
- Additional modes of breath sensing to complement the single contact probe, and a lighter needle insertion mechanism.
- Patient- and tool-specific system identification at the start of each procedure, which the authors recommend before any clinical deployment rather than relying on fixed plant parameters.
- Respiratory compensation under a remote center-of-motion constraint around the insertion site, which would allow insertions at more extreme angles than the current linear approach permits.
Target Audience
Medical robotics researchers and control engineers working on motion compensation and image-guided intervention will get the most from this paper, particularly those interested in estimator design for quasi-periodic physiological signals and in the practical details of tuning a filter that must act on a 300 ms delayed measurement. Interventional radiologists and clinical collaborators evaluating what autonomous CT-bore biopsy would actually require will find the procedural state machine and the anomaly-response table directly useful. Graduate students entering surgical robotics will benefit from the unusually complete account of how sensing specifications were derived from measured human data, and from the honest reporting of what did not work well — the torque-based insertion sensing and the phantom's deformation behavior.
Authors’ abstract
Currently, CT-guided lung biopsy is a highly manual procedure requiring patient breath-holding, iterative manual needle insertions, and repeated CT confirmation scans to accurately target a tumor. Automating this process within the CT bore--without relying on manual needle placement or patient breath-holding--has the potential to reduce procedure time, patient discomfort, and radiation exposure. In this work, we present an integrated sensing and control system capable of…
The full abstract on arXivRelated papers
- Bi-manual Stabilization of the Cervical Spine for Safe Physical Human-Robot Interaction in Rolling Maneuvers 2026-10-03 · cs.RO · Also by Ananya Rao, Derek Chen +2
- Reeling It In: Flexible Needle Pick Up via Thread Manipulation for Autonomous Suturing 2026-07-28 · cs.RO · Also by Emma Huang
- Physical Twins: Accelerating and Enabling Robot Learning with Phantom Platforms 2026-10-03 · cs.RO · Also by Derek Chen
- SurgFlow: 3D Object-Centric Contact Flow for Surgical Robot Manipulation 2026-09-27 · cs.RO · Also by Changwei Chen
- Reinforced Iterative Learning Control for elastic joint arms: Guaranteed cross-trajectory generalization under model uncertainties 2026-10-20 · cs.RO · Key terms of the title in common