> The first-generation Robotic Soft Esophagus (RoSE) could generate peristaltic waves only in open loop, because its conduit had no internal visibility and no embedded sensing, so there was no feedback…

**arXiv**: 2608.09602

**Published**: 2026-08-10

**Authors**: Dipankar Bhattacharya, Ryman Hashem, Leo K. Cheng, Weiliang Xu

## In plain words

The first-generation Robotic Soft Esophagus (RoSE) could generate peristaltic waves only in open loop, because its conduit had no internal visibility and no embedded sensing, so there was no feedback signal to model or control the device.

## AI summary

## Overview

**Research area:** Control of soft robotics for medical-device testing — specifically, data-driven modeling and closed-loop control of a soft robotic physical simulator of the human esophagus, used as an in vitro platform for esophageal stent testing.

**Technical level:** Advanced. The paper combines sparse nonlinear system identification (SINDYC), receding-horizon optimization, soft-lithography fabrication, and embedded sensing. Readers need working familiarity with state-space control and optimization to follow Sections IV–VI, though the problem framing is accessible to non-specialists.

**Scope:** This is the accepted-manuscript version of an article published in *IEEE Transactions on Industrial Electronics*, vol. 69, no. 10, pp. 10363–10373 (Oct. 2022; electronic publication 27 October 2021, doi: 10.1109/TIE.2021.3121755), and it describes how a sensor-equipped second-generation Robotic Soft Esophagus (RoSEv2.0) was modeled from data and then driven by a Sparse-Identification-of-Nonlinear-Dynamics-with-Control (SINDYC) model-based Model Predictive Control (MPC) scheme to produce prescribed peristaltic waves.

## What This Paper Is About

The first-generation Robotic Soft Esophagus (RoSE) could generate peristaltic waves only in open loop, because its conduit had no internal visibility and no embedded sensing, so there was no feedback signal to model or control the device. The authors rebuild the platform as RoSEv2.0 with an embedded array of Time-of-Flight (TOF) distance sensors and pressure sensors, learn discrete-time nonlinear models of the conduit directly from that sensor data, and embed those models inside an MPC loop so the robot can autonomously track a desired peristaltic wave profile and a desired chamber-pressure profile.

## Key Contributions

1. **A sensing-enabled soft esophagus (RoSEv2.0).** The outer casing of the original RoSE is replaced with a transparent PDMS (SYLGARD 182, Dow, USA) layer, onto which an array of VL6180X TOF sensors is bonded, giving lateral conduit-displacement measurements from outside the device — resolving the "no visibility inside the conduit" limitation of the original RoSE.
2. **Data-driven discrete-time models of the soft robot.** Discrete-time SINDYC (DTSINDYC) is applied to TOF and valve-pressure-sensor (VPS) data to discover discrete-time differential equation models consisting of constant, first-order and second-order polynomial terms in states and inputs. The authors state this avoids the derivative estimation from noisy data required by continuous-time SINDYC, and that a discretized model is needed for MPC anyway.
3. **First reported SINDYC-based MPC on a soft robot.** The authors claim that, to their knowledge, this SINDYC-based MPC framework has been applied for the first time in a soft robot, and that closed-loop MPC control has been implemented for the first time in RoSEv2.0.
4. **A sensor-agnostic modeling-and-control recipe.** The authors argue the methodology is not specific to a robot or a sensor, and demonstrate this by repeating the same pipeline with VPS data to control chamber air pressure instead of conduit displacement.

## Main Findings

- **Prescribed peristaltic waves were generated under MPC:** peristalsis waves of speed **20 mm·s⁻¹**, wavelength **75 mm**, and amplitudes **5, 7.5, and 10 mm** were successfully generated by the MPC.
- **Model structure:** both DTSINDYC models (M₁ from TOF data, M₂ from VPS data) use three state variables — displacement in mm for M₁, pressure in kPa for M₂ — under peristaltic control actuation.
- **Layer selection:** at least three RoSEv2.0 layers are needed to match human esophageal peristalsis wavelengths, so layers **L₅, L₆, and L₇** were chosen as the controlled states, selected with emphasis on TOF measurement accuracy.
- **Hardware and signal resolution:** the 12 layers each carry four air chambers, giving 12 independent inputs driven by ITV-0030-3BS (SMC, USA) electro-pneumatic valves rated 1–500 kPa over a 0–10 V input. With 8-bit DACs at 0.02 V·step⁻¹, each digital command corresponds to 1 kPa·step⁻¹. Full conduit occlusion requires a maximum of **47 kPa**, which is used as the upper control limit for MPC-generated signals.
- **Stent-migration testing setup:** RoSEv2.0 under MPC was used for stent migration testing with food boluses of various consistencies, and stent migration data for a candidate stent were recorded for various peristalsis speeds and bolus swallow conditions. **The quantitative stent-migration outcomes are not present in the available text**, which truncates partway through the Results section.
- **Tracking-quality metrics are not reported in the available text:** the truncated content does not include numerical tracking errors (e.g., RMSE or NRMSE values) for the MPC versus the reference peristaltic trajectory, nor model-fit statistics for M₁ and M₂.
- **Stability argument:** the authors state and prove Theorem 1 — if the plant model is unchanged and stable throughout control-law computation, and settling to the reference occurs within the prediction horizon, then MPC is closed-loop stable despite a finite horizon. The proof replaces the infinite-horizon tail of the cost with a terminal weight satisfying a discrete-time Lyapunov equation.
- **Timing design example:** with c = 20 mm/s, chamber spacing 15 mm, w = 120 mm and T_s = 0.1 s, the discrete delay between consecutive reference peaks is q = 7 time steps; each new control triple is right-shifted by three positions in an accumulator, and it takes 3q+1 iterations for a command to reach layers L₅–L₇. Layers L₁ and L₁₂ are not actuated because bolus feeder pipes occupy those locations.
- **A stated limitation of the scheme:** achieving higher peristalsis wave speeds becomes difficult if the MPC loop time T_s increases.

## Methodology in Plain English

The team first rebuilt the physical robot so it could be observed while it works. Each of the 12 layers has four air chambers; pressurizing a whorl of chambers closes the conduit at that layer, mimicking the circular muscle contraction of a real esophagus. A transparent silicone outer skin was bonded on, and small light-based distance sensors were glued to it to watch how far the wall moves outward at each layer. The whole body is cast from Ecoflex 0030 silicone with no rigid skeleton, and is driven by 12 pneumatic valves commanded from a Raspberry Pi 4 Model B through ADC/DAC boards. The TOF sensors were calibrated for range offset and crosstalk through the PDMS, and cross-checked against a Logitech C922 webcam using an OpenCV CSRT tracker with two 4 mm retroreflective markers.

Second, they ran the robot in open loop with many different pressure trajectories and recorded what the sensors saw. Third, they fed those state/input pairs into DTSINDYC, which builds a library of candidate terms — constants, linear states and inputs, and products of them — and uses sparse regression to keep only the handful of terms that actually explain how the state at the next time step follows from the current state and command. The result is a short, readable difference equation rather than a black-box neural network.

Fourth, they plugged that difference equation into an MPC controller. At each time step the controller looks N_p steps ahead, using the learned model to roll the state forward, and picks the sequence of valve commands that best keeps the predicted trajectory on a reference while penalizing large commands and large command changes, subject to the 0–47 kPa digital bounds. Only the first command is applied, then the whole calculation repeats one step later. The reference itself is a discretized, time-shifted cosine: three copies of the peristaltic wave profile, offset in time by the delay between adjacent layers, so that layers L₅, L₆, and L₇ reproduce a travelling wave of a chosen speed and wavelength. Commands are queued in a shift register so a wave generated at one layer propagates down the conduit at the right delay.

Finally, they ran stent-migration trials with boluses of different consistencies under this closed-loop peristalsis, and repeated the modeling-and-control pipeline on the pressure sensors instead of the displacement sensors to show the approach is not tied to one sensor type.

## Why This Matters

Stent migration — the stent slipping out of place as peristaltic waves push against it — is a recognized shortcoming for patients treated for malignant esophageal stricture, and testing stent designs in patients raises major ethical concerns, so evidence for one design over another has largely come from analytical and numerical studies. A soft robotic esophagus that can reproduce controllable, repeatable peristalsis gives researchers a physical testbed between simulation and clinical trial.

Real-world applications:

- **Stent design screening:** comparing self-expandable stent designs against realistic peristaltic loading before implantation, without patient involvement.
- **Food-bolus and dysphagia research:** studying how boluses of different consistencies interact with peristalsis and with an implanted stent.
- **Medical training and device validation:** a repeatable bench model of esophageal motility for training or for regulatory-style device characterization.
- **Transferable soft-robotics control:** the identification-then-MPC recipe applies to any soft actuator where input–output data can be collected, including soft gastric simulators and other pneumatically driven continuum devices.

**Industry relevance:** the work sits at the intersection of medical-device testing and soft-robotics control. Pneumatic soft actuators, embedded low-cost TOF sensing, and single-board-computer control are all commodity components, and the paper's argument that a sparse, interpretable data-driven model can replace a first-principles model — which the authors say is unobtainable for highly compliant silicone structures with effectively infinite degrees of freedom — is directly useful to anyone industrializing soft robots that must hit tight performance specifications.

## Future Directions

- **Publish the quantitative tracking and migration results.** The available text truncates before reporting trajectory-tracking error and stent-migration distances across speeds and bolus consistencies, which are needed to judge how well the closed loop actually performs.
- **Overcome the loop-rate ceiling.** The authors note that higher peristalsis speeds become difficult to achieve when the MPC cycle time T_s grows; reducing model or solver cost, or reformulating the optimization, is an obvious next step.
- **Scale the sensing beyond three layers.** Only layers L₅–L₇ are modeled and controlled, with L₁ and L₁₂ left unactuated to accommodate bolus feeder pipes; extending closed-loop control along the full conduit would allow more realistic, longer wave propagation.
- **Validate transferability more broadly.** The claim that the method is robot- and sensor-agnostic rests on a single second demonstration using pressure sensors on the same device; testing it on other soft robots and other sensor modalities would substantiate that claim.

## Target Audience

Researchers and graduate students in soft robotics and applied control who are interested in data-driven modeling and MPC for highly compliant, hard-to-model systems; engineers building physical simulators or testbeds for medical devices such as esophageal stents; and medical-device or gastroenterology researchers who need a controllable in vitro model of peristalsis. Readers looking for clinical outcomes or comparative stent-performance conclusions will not find them in the available text, which stops before the stent-migration results are presented.

## Authors’ abstract

Strictures caused by esophageal cancer can narrow down the esophageal lumen, leading to dysphagia. Palliation of dysphagia has driven the development of a Robotic Soft Esophagus (RoSE), which provides a novel in vitro platform for esophageal stent testing and food viscosity studies. In RoSE, peristaltic wave generation and control were done in an open-loop manner since the conduit lacked visibility and embedded sensing capability. Hence, in this work, RoSE version 2.0…

[The full abstract on arXiv](https://arxiv.org/abs/2608.09602)

[Read the original paper](https://arxiv.org/abs/2608.09602)

## Related papers

- [RoSE: A Robotic Soft Esophagus for Endoprosthetic Stent Testing](https://robotics.world/paper/rose-a-robotic-soft-esophagus-for-endoprosthetic-stent-testing-2608.09891) 2026-08-10 · cs.RO · Also by Dipankar Bhattacharya, Leo K. Cheng +1
- [Tri-Space Operational Control of Redundant Multilink and Hybrid Cable-Driven Parallel Robots Using an Iterative-Learning based Reactive Approach](https://robotics.world/paper/tri-space-operational-control-of-redundant-multilink-and-hybrid-cable-2607.29500) 2026-07-31 · cs.RO · Also by Dipankar Bhattacharya
- [Real-time Estimator of Actuator Control and Health (REACH) on an Eel-Inspired Soft Robot](https://robotics.world/paper/real-time-estimator-of-actuator-control-and-health-reach-on-an-eel-2608.14865) 2026-08-14 · cs.RO · Key terms of the title in common
- [Automated Straight-line Sewing of Stretchable Fabrics with Different Lengths](https://robotics.world/paper/automated-straight-line-sewing-of-stretchable-fabrics-with-different-2607.29464) 2026-07-31 · cs.RO · Also by Dipankar Bhattacharya
- [Real-Time Shape Control of Multi-Segment Soft Robotic Arms Using Koopman Operators with Global and Local Observables](https://robotics.world/paper/real-time-shape-control-of-multi-segment-soft-robotic-arms-using-2609.03175) 2026-09-02 · cs.RO · Key terms of the title in common

---

Canonical: https://robotics.world/paper/nonlinear-model-predictive-control-of-a-robotic-soft-esophagus-2608.09602