Modular Robotic Catheters for Endovascular Aneurysm Repair

Fenestrated and branched endovascular aneurysm repair (FEVAR/BEVAR) requires surgeons to steer catheters and guidewires into the renal, mesenteric, and iliac branches of the abdominal aorta so that…

The paper’s first page · arXiv:2607.25807
arXiv
2607.25807 · Preprint · not peer-reviewed
Published
2026-07-28
Authors
Alex Ranne, Jinshi Zhao, Ali Anil Demircali, Songli Moey, Ayhan Aktas, Burak Temelkuran, Nassir Navab, Ferdinando Rodriguez y Baena

In plain words

Fenestrated and branched endovascular aneurysm repair (FEVAR/BEVAR) requires surgeons to steer catheters and guidewires into the renal, mesenteric, and iliac branches of the abdominal aorta so that stent grafts can be deployed without cutting off blood flow to vital organs.

AI summary

Overview

Research area: Medical robotics and minimally invasive surgical instrumentation, specifically tendon-driven continuum (steerable) catheters for endovascular aortic repair.

Technical level: Intermediate. The paper assumes familiarity with catheter mechanics, tendon-driven actuation, and constant-curvature kinematics, but each component is described concretely enough for a reader with a general robotics background.

Scope: The paper presents and evaluates a 6 Fr, two-segment, four-degree-of-freedom tendon-driven robotic catheter — fabricated by thermal fiber drawing and laser micromachining — together with a modular 3D-printed actuation handle, validated through finite element simulation, benchtop bending tests, a ring obstacle course, and an in-vitro aortic phantom study.

What This Paper Is About

Fenestrated and branched endovascular aneurysm repair (FEVAR/BEVAR) requires surgeons to steer catheters and guidewires into the renal, mesenteric, and iliac branches of the abdominal aorta so that stent grafts can be deployed without cutting off blood flow to vital organs. Existing commercial instruments struggle with vessel tortuosity and unfavorable access angles, which lengthens procedures, increases radiation exposure, and lowers success rates in difficult cases. The authors' goal is a smaller, more dexterous, and more affordable robotic catheter system that can reach these target branches reliably.

Key Contributions

  1. A bespoke two-segment steerable catheter with 4 degrees of freedom. The device is a 6 Fr (2 mm) polycarbonate body produced by thermal fiber drawing, with eight 200 µm tendon channels and a 1.1 mm central guidewire channel. Each of the two segments contributes two degrees of freedom, giving more dexterity than the single-segment steerable catheters the paper compares against.

  2. A scalable fabrication and assembly pipeline. Thermal drawing produces tens of meters of catheter tubing per run from a 40 mm diameter, 15 cm long 3D-printed preform; femtosecond laser micromachining then creates 0.2 mm by 0.8 mm tendon windows and offset stiffness-reducing profiles at different depths on each segment (400 µm distal, 200 µm proximal).

  3. An expandable, modular, 3D-printed actuation platform. The handle comprises a cap module, identical actuation modules, and an optional joystick module. Each actuation module uses one Dynamixel XL-330-M288-T driving a 0.5 MOD pinion meshed with two rack gears (6 cm travel), actuating an antagonistic tendon pair; dovetail joints let modules attach in four orientations, so the platform scales to 2 × n tendons for n actuators.

  4. A forward kinematic model and staged experimental evaluation. A constant-curvature model accounts for the 45° rotational offset between the two segments, and the system is assessed via COMSOL finite element analysis, benchtop bending characterization, a ring navigation course, and cannulation of an abdominal aortic phantom.

Main Findings

  • Profile depth controls stiffness predictably: In COMSOL simulations of a 12 cm polycarbonate tube (Young's modulus 2 GPa) with profile depths swept from 0 mm to 0.6 mm, bending stiffness decreased monotonically with depth, allowing a reduction of up to 83%. The authors nonetheless chose 0.4 mm for the distal segment and 0.2 mm for the proximal segment, because the proximal segment may need to support stent grafts or stiff guidewires.
  • Bending angles are competitive with target anatomy: The second segment alone reached a maximum bending angle of 35°, and both segments actuated together reached 72°. For multi-plane bending, the second segment bent up to 36° relative to the first segment while the first segment bent up to 38° to the left. The paper cites average vessel angles of 64.1° for the renal arteries, 51.5° for the superior mesenteric artery, and 40.5° for the iliac arteries, and argues the prototype is fit to traverse all of them.
  • Navigation was demonstrated but not quantified: The catheter traversed a three-ring 3D-printed PLA obstacle course (rings 10 cm apart, the two outer rings 15 cm tall, the middle ring 3 cm shorter and offset 2.5 cm from the centerline) and reached both renal arteries and the inferior mesenteric artery in a silicone abdominal aorta phantom (Elastrat, Geneva, Switzerland) while delivering colored dye. The paper reports these outcomes as image sequences rather than as success rates, completion times, or force measurements.
  • Two named failure modes remain: The catheter exhibits the "muscling" phenomenon, where the whole shaft moves when tendons are pulled, and the "curve alignment" phenomenon, an undesired rotation of the full catheter. The authors point to helical tendon channels, as used by Bogusky et al. and Abdelaziz et al., as the established remedy they have not yet adopted.
  • Friction is absent from the model: The finite element study explicitly neglects friction, which the authors identify as a gap that must be closed before clinical deployment.

Methodology in Plain English

The team started with a macroscopic 3D-printed preform — a 40 mm wide, 15 cm long polycarbonate block with the desired channel layout — and heated it in a three-zone furnace (120 °C, 190 °C, 85 °C) while pulling the softened material downward at 0.4 mm/s, with a preform downfeed of 1 mm/min. This necks the preform down to a 2 mm fiber while preserving the internal geometry; the fiber diameter follows from conservation of mass.

They then hand-fed 80 µm molybdenum wires into the tendon channels and used a femtosecond laser system (stated precision below 3 µm) mounted with a custom catheter rotation/translation stage to cut small windows exposing the wires, plus offset slot profiles that locally soften the tube. Tendons pulled out through the windows were clamped between two layers of wound 50 µm stainless steel wire, and the whole catheter was sleeved in 127 µm wall medical heat shrink.

The handle is entirely 3D-printed in PLA except fasteners and ferrules. Opposing tendons are driven by a single motor through a rack-and-pinion, halving the actuator count relative to a fully independent scheme; four such modules give the four degrees of freedom. A joystick module clamps to the handle so one hand can hold and steer the device.

For evaluation, the authors simulated bending stiffness with beam theory (EI = WL³/3δ), measured bending angles on an optical table, drove the catheter through a ring course using a linear stage for insertion and an Xbox controller for steering, then repeated the setup in a silicone aortic phantom to cannulate the renal and mesenteric branches.

Why This Matters

Impact on research: The paper contributes a fabrication route — thermal drawing plus laser micromachining — that produces multi-lumen continuum robots at scale and low cost, and pairs it with a handle architecture that scales to arbitrary even numbers of antagonistically paired tendons. It also provides an explicit stiffness-versus-profile-depth trade-off analysis that other groups designing laser-profiled catheters can reuse when choosing between dexterity and the ability to support stent grafts.

Real-world applications:

  • FEVAR/BEVAR target vessel catheterization, particularly the renal and mesenteric branches that the cited 2021 Tschischka et al. study found difficult even with multiple access points.
  • Peripheral and cardiac interventions requiring selective access, mirroring the domains of the commercial steerable catheters the paper surveys (2.4 Fr to 12 Fr).
  • Neurovascular and pediatric procedures, where the paper cites the 2.4 Fr SwiftNINJA and 2.5 Fr Benfit systems as evidence that small diameter plus a working channel is clinically valuable.
  • Thrombolysis and contrast/dye delivery, which the in-vitro study mimics by injecting colored dye into three arteries.

Industry relevance: The paper positions itself directly against the Magellan system (Hansen Medical, FDA approval 2012), whose adoption the authors attribute to cost, catheter size, and setup time rather than capability. A 3D-printed handle with hobby-grade Dynamixel servos and a fiber-drawn disposable catheter body is an explicit argument that the cost and size barriers are surmountable. The cited 83% success rate in challenging FEVAR/BEVAR cases versus the standard figure of over 95% frames the clinical market need.

Future Directions

  1. Model and compensate friction. The finite element study neglects friction; the authors plan feedforward models to compensate for tendon friction and sheath interaction forces.
  2. Eliminate muscling and curve alignment. The paper proposes a helical-structured passive section to counter undesired shaft motion and rotation induced by tip actuation.
  3. Explore softer materials. Design optimization using elastomers instead of polycarbonate is named as a next step for fabrication.
  4. Add sensing and autonomy. The authors target shape and proximity sensing, more autonomous navigation to reduce contact forces from manual operation, and integration with other surgical devices.

Target Audience

Surgical robotics researchers working on continuum and tendon-driven mechanisms; medical device engineers interested in scalable catheter fabrication, laser profiling, or low-cost modular actuation; and interventional clinicians or clinical collaborators who want an accessible account of what a two-segment steerable catheter can and cannot yet do in aortic anatomy. Readers seeking quantified clinical performance data — success rates, procedure times, or force measurements in the phantom and ring studies — will find those are not reported here.

Authors’ abstract

Fenestrated/Branched endovascular aneurysm repair (FEVAR/BEVAR) require surgeons to navigate catheters and guidewires into various branches of the abdominal aorta, before deploying stent grafts to alleviate pressure on the aneurysm. Previous clinical studies suggests that surgeons continue to struggle with vessel access using standard commercial instruments, prolonging the procedural time and inducing further complications. In this work, we present two contributions to…

The full abstract on arXiv
Read the original paper