Selectively Actuated Fiberbots for High-precision Minimally Invasive Surgery
Long, thin instruments are needed to reach deep anatomical targets through endoscope channels and tortuous pathways, but controlling the tip precisely becomes harder as the instrument gets longer…

- arXiv
- 2610.12636 · Preprint · not peer-reviewed
- Published
- 2026-10-08
- Authors
- Jinshi Zhao, Libaihe Tian, Alex Ranne, Qindong Zheng, Yue Li, Xiaotong Guo, Ziqi Yang, Ali Anil Demircali, Mohamed E. M. K. Abdelaziz, Bruno Gil Rosa, Ferdinando Rodriguez y Baena, Burak Temelkuran
In plain words
Long, thin instruments are needed to reach deep anatomical targets through endoscope channels and tortuous pathways, but controlling the tip precisely becomes harder as the instrument gets longer, because shaft deformation, friction, backlash, and channel interactions blur the proximal-to-distal relationship.
AI summary
Overview
Research area: Medical robotics — soft/continuum surgical instruments, specifically thermally drawn electrothermally actuated fiber robots ("fiberbots") for minimally invasive surgery. Technical level: Advanced (assumes familiarity with soft robotics, thermal drawing fabrication, electrothermal actuation, and surgical robotics platforms). Scope: The paper presents a 1.9 mm × 1.2 m thermally drawn fiber robot whose Joule heating is confined to a 10 cm distal segment, characterizes its motion and thermal behavior, and demonstrates a da Vinci–integrated phantom scanning task.
What This Paper Is About
Long, thin instruments are needed to reach deep anatomical targets through endoscope channels and tortuous pathways, but controlling the tip precisely becomes harder as the instrument gets longer, because shaft deformation, friction, backlash, and channel interactions blur the proximal-to-distal relationship. Electrothermally actuated fiber robots solve part of this by putting the actuator at the tip, but earlier versions had to power the entire fiber length, wasting energy as heat along mechanically passive shaft. The goal here is a fiberbot that heats and actuates only its distal segment, while remaining small enough to pass through an off-the-shelf endoscope working channel and mechanically uncoupled from it.
Key Contributions
- A selective actuation architecture for fiberbots. Three "copper–stainless steel–copper" wire triplets are embedded in nine side lumens; the high-resistivity stainless steel (6.9 × 10⁻⁷ Ω·m, roughly forty times copper's 1.68 × 10⁻⁸ Ω·m) is electrically connected only across a 10 cm distal segment, so power concentrates there while the passive shaft transmits current through low-resistance copper.
- A thermal-drawing fabrication route for a multi-lumen, wire-embedded surgical fiber. A 3D-printed polycarbonate preform (15 cm long, 4 cm diameter) with nine side lumens is drawn with wires fed from a nine-spool platform, producing a 1.9 mm outer diameter / 1 mm inner diameter fiber that houses a central channel for commercial surgical laser fibers.
- A complete device and control stack. Proximal electrical access via femtosecond laser micro-machined 3 mm × 1 mm windows, distal laser spot-welded connections, a 3D-printed PLA handle (50 mm × 30 mm ellipse, 10 cm long), and a six-op-amp buffer circuit driven through a cRIO-9025 real-time controller in LabVIEW.
- Quantified motion, thermal, and power benefits, plus a system-level demonstration. Sub-100 μm open-loop path-following error, thermal confirmation that the passive section stays near ambient, an ~89% power reduction versus the previous architecture at 1.0 m passive length, and a da Vinci–coupled endoluminal phantom scanning task.
Main Findings
- Precision below 100 μm: Path following of a star and a spiral shape with an open-loop controller produced a maximum positioning error below 100 μm.
- Linear input–displacement behavior: Across 500 random voltage sets applied to two of three wire pairs at a time, displacement scaled approximately with electrical power (and therefore with V²) along each actuation direction, which is what made the open-loop power-based controller viable.
- Selective heating confirmed by thermal camera: In a side view spanning active and passive sections, the active section rose by approximately 28.5 °C while the passive section rose by only around 1.4 °C.
- Large power savings that grow with length: For a fixed 0.10 m active section producing a 2 mm tip displacement, required power rose only from about 0.40 W to 0.75 W as passive length went from 0 to 1.0 m, versus roughly 6.6 W for the previous architecture at 1.0 m — about 11% of the power, an approximately 89% reduction.
- Temperature and displacement both scale with V²: In the single-wire-pair power sweep (0.1 W steps, 20 s holds), tip displacement and temperature rise increased approximately proportionally to the square of applied voltage.
- Clinical-workflow feasibility: With three da Vinci instruments used concurrently (camera, grasper, fiberbot-plus-instrument assembly), an operator introduced the fiber into an endoluminal phantom, exposed a black target region, and scanned the target boundary with a joystick in approximately 4 minutes.
- The passive segment is not free in the old design: The analysis shows that mechanically clamping a heated section stops its motion but does not stop its power consumption — the core motivation for localizing the heater.
Methodology in Plain English
The team built the fiber in two stages. First they 3D-printed a large polycarbonate "preform" — a scaled-up version of the final cross-section, with a central hole and nine small side holes — dried it in a vacuum oven for two days, mounted it on a PEEK rod with nine spools of wire feeding through the holes, and lowered it through a three-zone furnace at 120 °C / 190 °C / 85 °C. Above polycarbonate's glass transition (~112 °C) the plastic softens enough to be pulled down to fiber scale on a rotating capstan, with the wires drawn in and held in place.
To make the actuation selective, they wired the stainless steel wire to one copper wire at the tip (pulled out, twisted, laser spot-welded at 120 V, and covered with Kapton tape) and to the other copper wire partway up the shaft, accessed through laser-cut windows on the surface. Applying voltage to the two copper wires at the proximal end then drives current through only the short stainless steel span, which heats and bends.
For characterization they used a 10 cm active fiberbot with a single-mode optical fiber down its center channel lit by a red LED, tracking the glowing tip with a digital microscope. They applied 500 random voltage sets (held 20 s, with 40 s of zero voltage between sets), decomposed each resulting tip displacement along the two actuation directions, and fit the voltage–power–displacement relationship. They then inverted that relationship to steer along star and spiral paths, applying each computed voltage for 100 ms. A thermal camera, first with a 2× close-up lens and then with a wide-field lens for side views, measured cross-sectional and along-length temperature during stepped power increases. The power comparison against the prior design was a calculation summing wire-segment resistances along the current path, neglecting contact resistance and temperature dependence of resistivity. Finally, they mounted the fiber in a bespoke holder on a da Vinci instrument and ran an abdominal phantom task with a red LED standing in for a laser.
Why This Matters
Impact on research: The paper reframes a scaling limitation of electrothermally actuated fiber robots — that heating the whole shaft wastes power and creates unwanted coupling — as a design problem solvable by material and wiring geometry rather than by more power. It gives the field a concrete, fabricable architecture (thermal drawing plus selective electrical interconnection) and a characterization baseline (sub-100 μm open-loop error, measured active/passive thermal split) that future fiberbot designs can be compared against.
Real-world applications:
- Laser delivery and ablation in endoluminal and laparoscopic procedures, where the fiber's 1 mm central channel accepts commercial surgical laser fibers.
- Precise scanning or painting of a lesion or tumor boundary at the surgical site, as simulated in the phantom study.
- Catheter-based and endoscopic interventions requiring dexterous distal motion through long, tortuous pathways.
- Use as a add-on dexterous tip inside existing robotic platforms such as the da Vinci system, without replacing the platform.
Industry relevance: The device is explicitly designed around off-the-shelf endoscope working channels and mechanically uncoupled from the endoscope, and it is shown interfacing with an existing commercial surgical robot. That framing targets retrofittable, disposable-scale instruments rather than new capital equipment. The thermal drawing process is also presented as a lower-tooling-cost alternative to extrusion and molding, which matters for manufacturability of high-aspect-ratio multi-lumen devices.
Future Directions
- Higher glass-transition-temperature polymers for the fiber body, to allow more power delivery and a larger motion range, with the safety implications of higher temperatures addressed.
- Replacing the Kapton tape tip protection with a polymer enclosure, to guarantee reliable electrical insulation in liquid environments — an explicit prerequisite for wet clinical use.
- Closing the loop: the demonstrated control is open loop with sub-100 μm error; adding distal sensing could reduce error further and compensate for the unmodeled effects (contact resistance, resistivity temperature dependence) that the power analysis deliberately neglected.
- In vivo and full-system validation: the current evidence is a 10 cm active-section bench characterization plus a phantom study, so animal or clinical work and testing with longer active sections and real tissue interaction remain open.
Target Audience
Robotics and medical-device researchers working on soft continuum manipulators, catheter and endoscopic robotics, and thermally drawn or fiber-based devices; surgical robotics engineers evaluating add-on distal dexterity tools; and translational clinicians or industry developers interested in how a sub-millimeter-precision fiber actuator might be integrated into existing laparoscopic and endoscopic workflows. Readers without a background in electrothermal actuation or polymer fiber fabrication will find the methods sections dense, though the motivation and results are accessible.
Authors’ abstract
Long, slender instruments are essential for minimally invasive surgery (MIS), but achieving precise distal manipulation becomes increasingly difficult when actuation extends along the entire instrument. Here, we present an electrothermally actuated fiber-based robot designed for MIS procedures, in which Joule heating is concentrated in the distal active segment. The fiber is fabricated using a thermal drawing process and uses embedded metallic wires for thermal actuation.…
The full abstract on arXivRelated papers
- Modular Robotic Catheters for Endovascular Aneurysm Repair 2026-07-28 · cs.RO · Also by Alex Ranne, Ali Anil Demircali +3
- Forbidden Region Dynamic Active Constraints in Robot-Assisted Minimally Invasive Surgery 2026-08-04 · cs.RO · Also by Ferdinando Rodriguez y Baena
- LocoWM: High-Precision Locomotion through World-Model-Guided Residual Adaptation 2026-09-30 · cs.RO · Key terms of the title in common
- High-Precision RCM Motion for Ophthalmic Robotics via Hierarchical QP Control and Authority-Guided Recovery Planning 2026-10-20 · cs.RO · Key terms of the title in common
- ActGaze: Learning Action-Grounded Gaze through Counterfactual Visual Interventions for High-Precision Manipulation 2026-09-24 · cs.RO · Key terms of the title in common